Product Details
Name:
S1100 ICU VENTILATOR \ Mechanical Ventilation In Icu
Ventilation Mode:
PRVC, APRV, DUOLEVEL, V-SIMV, P-SIMV, P-A/C, V-A/C,
Pipeline:
High Temperature Resistance
Function:
Self-check\Multiple Ventilation Modes
Screen:
15" TFT Touch Screen
Voltage:
220 V ± 22 V
Main Engine:
560 * 560 * 615 Mm
Air Compressor:
685 * 690 * 1140 Mm
Payment & Shipping Terms
Product Description
S1200 ICU VENTILATOR Icu Oxygen Machine
 
Application
 
The ICU oxygen machine makes a good performance in operation room, ICU department and emergency treatment.
It used to assist or replace the sponta-neous breathing for adult, pediatric and neonatal patient.
23 years of experience in Market-oriented ventilator make us professional and reliable, satisfy all your need in ventilation. Due to the flflexible confifiguration, good quality and competitive price, the S1200 has soon become the superstar of market.
 
Features
 
  1. Multiple ventilation modes, suitable for ICU, emergency department and operation room etc: (PRVC, APRV, DUOLEVEL, IPPV, PCV, PSV,
  2. V-SIMV, P-SIMV,V-A/C, P-A/C, SPONT/CPAP, SIGH, MANUAL)
  3. Pressure and flow dual triggering mechanism make patient brea the better and more comfortable.
  4. Rapid oxygen supply, automatically offer high flow rate oxygen within two minutes.
  5. Pressure-Time, Flow-Time, Volume-Time, ETCO2 -Time oscillogram, real-time display the condition of patient in all respects.
  6. Separate design of electronic circuit and gas flow rate keep safe running of ventilator.
  7. Compact long life internal battery can provide emergency power, avoid risk of patient.
  8. Three level alarm system display overall condition of machine, easier for users to find and solve problems according to indication.
  9. Oscillogram freezing and recovery offer doctors a better understanding of the real-time situation for patient.
  10. Self-check before operation, ensure every port of machine is in good condition.

 

Safety
 
  1. 14 types of sound and visual alarm information, easier for users to do some error checking and troubleshooting.
  2. Built-in oxygen concentration sensor, ensure stable precision of oxygen concentration.
  3. Easy to move with four casters, easy to stop with two brakes.
  4. Separate design of electronic circuit and gas flow rate keep safe running of ventilator.
  5. Compact long life internal battery can provide emergency power, avoid risk of patient.
  6. Self-check before operation, eliminate system mistake.
Friendly interface
15" TFT LCD screen display screen saver mode.
14 types of alarm and protection function.
3 of 5 oscillographs can be displayed on the screen at the same time.
5 oscillograms for your choice, 3 of them can be displayed on the screen at the same time.
Emergency Department ICU Oxygen Machine Nurse Ventilator S1200 0
Air compressor
Automatic detection and switch of central gas, cylinder gas supply or air compressor.
Make low noise, create quiet environment for doctors and patients.
Longer service life and waterproof material.
Digital show for parameters, easier and clearer for users know real-time condition.
Emergency Department ICU Oxygen Machine Nurse Ventilator S1200 1
Caster
Easy to move with 4 casters,easy to stop with 2 brakes.
Humidifier
Heat and wet breathing gas, makes it more comfortable for patient to
breathe in.
3 types of humidifier for your choices: one with thermometer for
standard, one with digital shown for option, another one has have dual
pipeline heating function and servo control function for option.
Emergency Department ICU Oxygen Machine Nurse Ventilator S1200 2
Nebulization function
Make medicine into small liquid and particle, easier and quicker for patient to breathe in.
Oxygen sensor
Built-in oxygen concentration sensor, ensure stable precision of oxygen concentration.
Breathing circuit
High temperature resistance breathing piplines
is reusable and anti-polution.
Emergency Department ICU Oxygen Machine Nurse Ventilator S1200 3
Ventilation mode
PRVC, APRV, DUOLEVEL, IPPV, PCV, PSV, V-SIMV,
P-SIMV, V-A/C, P-A/C, SIGH, SPONT/CPAP, MANUAL
 
   
Ventilator parameter range
Tidal volume (Vt) 0, 20 ~ 2000 mL
Frequency (Freq) 1 min ~ 100 min
Oxygen concentration 21 % ~ 100 %
I:E 4 : 1 ~ 1 : 8
PEEP 0 cmH2O ~ 40 cmH2O
Pressure triggering sensitivity (Ptr) -20 cmH2O ~ 1000 cmH2O (Based on PEEP)
Monitoring parameter
Frequency (Freq) 0/min ~ 100/min
Tidal volume (Vt) 0 mL ~ 2500 mL
MV 0 L/min ~ 99 L/min
Airway pressure 0 cmH2O ~ 100 cmH2O
Dynamic lung compliance monitoring 1 mL/cmH2O ~ 1000 mL/cmH2O
Oxygen concentration 15 % ~ 100 %

 

Alarm and protection
AC power failure alarm Power failure or no connection
Internal backup battery low voltage alarm ≤ 11.3 ± 0.3 V
No tidal volume No tidal volume within 6 s
High minute volume alarm 5 L/min ~ 99 L/min
Low minute volume alarm 1 L/min ~ 30 L/min
High airway pressure alarm 20 cmH2O ~ 100 cmH2O
Low airway pressure alarm 0 cmH2O ~ 20 cmH2O
High oxygen concentration alarm 19 % ~ 100 %
Low oxygen concentration alarm 18 % ~ 99 %
Continuous pressure alarm (PEEP + 1.5 cmH2O) over 16s
Suffocation warning 5 ~ 60 s
Fan error Show on screen
Oxygen deficit Show on screen
The maximum limited pressure < 12.5 kPa
   
Working condition
Gas source O2, Air
Pressure 280 kPa - 600 kPa
Voltage 220 V ± 22 V
Power frequency 50 Hz ± 1 Hz
Input power 900 VA(With air compressor)
250 VA(Without air compressor)
Oscillogram Wooden case packing size
P-T (Pressure-Time) Main engine : 560 * 560 * 615 mm
F-T (Flow-Time) G.W. : 30 KG N.W. : 15 KG
V-T (Volume-Time) Air compressor : 685 * 690 * 1140 mm
ETCO2 -T (End-Tidal CO2 -Time) G.W. : 138 KG N.W. : 63 KG
P-V Loop (Pressure-Volume Loop)  
F-V Loop (Flow-Volume Loop)  
P-F Loop (Pressure-Flow Loop)  

Emergency Department ICU Oxygen Machine Nurse Ventilator S1200 4Emergency Department ICU Oxygen Machine Nurse Ventilator S1200 5Emergency Department ICU Oxygen Machine Nurse Ventilator S1200 6