Ventilation mode |
PRVC, APRV, DUOLEVEL, V-SIMV, P-SIMV, P-A/C, V-A/C, |
IPPV, PCV, PSV, SPONT/CPAP, SIGH, MANUAL |
Alarm and protection | |
AC power failure alarm | Power failure or no connection |
Internal backup battery low voltage alarm | ≤ 11.3 ± 0.3 V |
No tidal volume | No tidal volume within 6 s |
High minute volume alarm | 5 L/min ~ 99 L/min |
Low minute volume alarm | 1 L/min ~ 30 L/min |
High airway pressure alarm | 20 cmH2O ~ 100 cmH2O |
Low airway pressure alarm | 0 cmH2O ~ 20 cmH2O |
High oxygen concentration alarm | 19 % ~ 100 % |
Low oxygen concentration alarm | 18 % ~ 99 % |
Continuous pressure alarm | (PEEP + 1.5 cmH2O) over 16s |
Suffocation warning | 5 ~ 60 s |
Fan error | Show on screen |
Oxygen deficit | Show on screen |
The maximum limited pressure | < 12.5 kPa |
Working condition | |
Gas source | O2, Air |
Pressure | 280 kPa - 600 kPa |
Voltage | 220 V ± 22 V |
Power frequency | 50 Hz ± 1 Hz |
Input power | 900 VA(With air compressor) 250 VA(Without air compressor) |
Wooden case size | |
Main engine | L 560 * W 560 * H 615 mm |
G.W.: 32 KG | N.W.: 19 KG |
Air compressor | L 685 * W 690 * H 1140 mm |
G.W.: 98KG | N.W.: 63 KG |