Friendly interface 15" TFT LCD screen display screen saver mode. 14 types of alarm and protection function. 3 of 5 oscillographs can be displayed on the screen at the same time. 2 minutes fast oxygen supply. |
|
Humidifier Heat and wet breathing gas, makes it more comfortable for patient to breathe in. 3 types of humidifier for your choices: one with thermometer for standard, one with digital shown for option, another one has have dual pipeline heating function and servo control function for option. |
|
Air compressor Automatic detection and switch of central gas, cylinder gas supply or air compressor. Make low noise, create quiet environment for doctors and patients. Longer service life and waterproof material. Digital show for parameters, easier and clearer for users know real-time condition. |
|
Caster Easy to move with 4 casters,easy to stop with 2 brakes. |
|
Breathing circuit High temperature resistance breathing piplines is reusable and anti-polution. Oxygen concentration sensor |
Ventilation mode | |
PRVC, APRV, DUOLEVEL, V-SIMV, P-SIMV, P-A/C, V-A/C, | |
IPPV, PCV, PSV, SPONT/CPAP, SIGH, MANUAL | |
Ventilator parameter range | |
Tidal volume (Vt) | 0, 20 ~ 2000 mL |
Frequency (Freq) | 1 min ~ 100 min |
Oxygen concentration | 21 % ~ 100 % |
I:E | 4 : 1 ~ 1 : 8 |
PEEP | 0 cmH2O ~ 40 cmH2O |
Pressure triggering sensitivity (Ptr) | -20 cmH2O ~ 0 cmH2O (Based on PEEP) |
Flow trigger sensitivity (Ftr) | 0.5 L/min ~ 30 L/min |
Pressure control (PC) | 5 cmH2O ~ 80 cmH2O |
Pressure support (PS) | 0 cmH2O ~ 80 cmH2O |
SIGH | 0 (off) 1/100 ~ 5/100 |
Apnea ventilation | OFF, 5 s ~ 60 s |
Pressure limit | 20 cmH2O ~ 100 cmH2O |
Monitoring parameter | |
Frequency (Freq) | 0/min ~ 100/min |
Tidal volume (Vt) | 0 mL ~ 2500 mL |
MV | 0 L/min ~ 99 L/min |
Airway pressure | 0 cmH2O ~ 100 cmH2O |
Dynamic lung compliance monitoring | 1 mL/cmH2O ~ 1000 mL/cmH2O |
Oxygen concentration | 15 % ~ 100 % |
Alarm and protection | |
AC power failure alarm | Power failure or no connection |
Internal backup battery low voltage alarm | ≤ 11.3 ± 0.3 V |
No tidal volume | No tidal volume within 6 s |
High minute volume alarm | 5 L/min ~ 99 L/min |
Low minute volume alarm | 1 L/min ~ 30 L/min |
High airway pressure alarm | 20 cmH2O ~ 100 cmH2O |
Low airway pressure alarm | 0 cmH2O ~ 20 cmH2O |
High oxygen concentration alarm | 19 % ~ 100 % |
Low oxygen concentration alarm | 18 % ~ 99 % |
Continuous pressure alarm | (PEEP + 1.5 cmH2O) over 16s |
Suffocation warning | 5 ~ 60 s |
Fan error | Show on screen |
Oxygen deficit | Show on screen |
The maximum limited pressure | < 12.5 kPa |
Working condition | |
Gas source | O2, Air |
Pressure | 280 kPa - 600 kPa |
Voltage | 220 V ± 22 V |
Power frequency | 50 Hz ± 1 Hz |
Input power | 900 VA(With air compressor) 250 VA(Without air compressor) |